Medical Device alert: CDSCO issues alert on MitraClip Clip Delivery System

This change in arm angle after deployment can be identified through fluoroscopy, said the Medical Device Alert.

561
Medical Device
Medical Device

Table of Contents

Last Updated on January 9, 2024 by The Health Master

Medical Device alert

The Central Drugs Standard Control Organisation (CDSCO) has issued a medical device alert on US-based healthcare and medical devices firm Abbott’s heart valve repair device, MitraClip Clip Delivery System, which is intended to treat a heart condition that causes the mitral valve on the left side of the heart not to close properly.

The alert was issued following an increased rate of reports of clips failing to establish final arm angles (EFAA) and for events of clip opening while locked (COWL), for which the manufacturer alerted the US drug regulator in September, 2022.

The EFAA is a procedural step where the user intentionally attempts to open a locked clip to verify that the locking mechanism is engaged.

This occurs during device preparation and prior to clip deployment. An EFAA failure occurs if the Clip opens during this verification step.

The COWL describes when the Clip Arm Angle increases post-deployment. In such cases, users observe a sloppage in the lock, resulting in an Arm angle greater than 10 degrees from the angle observed at deployment.

This change in arm angle after deployment can be identified through fluoroscopy, said the Medical Device Alert.

When the clip is locked, the locking mechanism is designed to more fully engage as the grippers push down on the clip Arms. During a COWL event, the lock slips as the Clip Arms open by a given degree, it added.

Any person with mitral regurgitation (MR) – a disorder in which the mitral valve on the left side of the heart does not close properly and leak occurs from the valve since it does not close all the way – who has to be treated with the use of MitraClip G4 Clip Delivery System and MitraClip NTR/XTR Clip Delivery System are expected to be affected with the malfunction, said the alert.

However, the CDSCO noted that it has not received any complaints from the market on this issue.

The US Food and Drug Administration (USFDA) has alerted the healthcare providers about potential clip lock malfunctions with the MitraClip Clip Delivery Systems manufactured by Abbott on September 8, 2022.

On September 8, 2022, Abbott issued an Urgent Medical Device Correction to inform health care providers about the issue, said the USFDA alert.

“An increased rate of reports of clip lock malfunctions has been observed before and after clip deployment. These events appear to occur in approximately 1.3% of MitraClip procedures and have been observed with all device models,” it said.

The potential risk to patients in the event of a clip lock malfunction includes ineffective treatment of MR and the potential need for additional interventions contributing to increased procedural risks such as bleeding, complications with implanting additional clips, and longer procedural times.

The majority of reported clip lock malfunction events have not been associated with adverse patient outcomes.

“Based on the available data on clip lock malfunctions and the associated risks, the USFDA believes that the probable benefits of the MitraClip device continue to outweigh the probable risks for the approved indications for use.

The USFDA is issuing this letter to ensure you are aware of the manufacturer’s recall notice and recommended actions,” it added.

The US drug regulator further added that MitraClip was first approved in 2013 to reduce MR in selected patients whose significant symptomatic MR and heart failure symptoms result from abnormalities of the mitral valve (commonly known as primary or degenerative MR); and whose risks for mitral valve surgery are prohibitive.

In 2019, a new indication for the device was approved to include treatment of patients with structurally normal mitral valves who develop heart failure symptoms and moderate-to-severe or severe MR due to left heart enlargement and diminished function (commonly known as secondary or functional MR) despite treatment with optimal medical therapy.

MitraClip has been the only percutaneous (implanted through the skin without open surgery) repair device approved in the United States to treat patients with MR, it said during the time.

Medical device alert: CDSCO issues alert on batches of peripheral stent from US-based Cordis

Medical Device Alert: CDSCO issues alert on these Medical Devices

Medical Device Alert: Urgent Medical Device Correction issued regarding this Possible Risk

Medical Device Alert: CDSCO issues an alert on cyber security risks

Medical Device Alert: CDSCO issues alert on these insulin pumps: Medical Devices

Medical Device Alert: DCGI issues safety alerts on Medical Devices

Drug Alert: NSQ Drugs Lists

Medical Device Alert: Lists

Safety alerts issued by IPC

Drug recall: Vancomycin Hydrochloride Injection recalled due to this reason

NPPA is in the process of fixing ceiling prices of 63 formulations

Indian scientists discover new Anti-Cancer Agent

Anti-dumping investigation begins on import of Vitamin-A Palmitate

CDSCO panel gives nod for BE study of Tadalafil Oral Jelly

List of approved New Drugs Updated

No need to recall, re-label the drug for price revision if…….:NPPA

ICMR releases Antimicrobial guidelines 2022

Registration for sale of Medical Devices: Procedure and conditions

Latest Notifications regarding Pharmaceuticals

For informative videos by The Health Master, click on the below YouTube icon:

YouTube Icon

For informative videos on Medical Store / Pharmacy, click on the below YouTube icon:

YouTube Icon

For informative videos on the news regarding Pharma / Medical Devices / Cosmetics / Homoeopathy etc., click on the below YouTube icon:

YouTube Icon

For informative videos on consumer awareness, click on the below YouTube icon:

YouTube Icon
Telegram
WhatsApp
Facebook
LinkedIn
YouTube Icon
Google-news